Electromagnetic flow converter for combination with OPTIFLUX and WATERFLUX flow sensors. KROHNE 03/ – – TD IFC R05 en. The documentation is only complete when used in combination with the relevant documentation for the . 2. The high-performance solution. IFC is the only electromagnetic flow converter with 3 x diagnostics of instrument, application and .

Author: Bashura Zulusida
Country: Niger
Language: English (Spanish)
Genre: Career
Published (Last): 27 July 2004
Pages: 219
PDF File Size: 18.14 Mb
ePub File Size: 10.36 Mb
ISBN: 459-6-14802-607-2
Downloads: 95599
Price: Free* [*Free Regsitration Required]
Uploader: JoJojora

Simulation of the flow fraction for partly filled pipes. Slide the display and jrohne the metal puller devices back into the housing.

If “standard” is set, set field frequency in Fct.

After that the device immediately begins measuring, and the current values are displayed. If you require data that is more relevant to your specific application, please contact us or your local representative.

I agree to the terms and privacy policy.

Krohne ifc Manuals

Backplane invalid The data record on the backplane is invalid. For further information refer to Mounting plate, wall-mounted housing on page Signal Cable B type BtsConstruction Page 36 Setting as indicated on ifd nameplate C 1. Our Products Vissers Sales Corp. Depending on the device version, not all functions are available. Information Concerning The Documentation Deposits in measuring tube or electrode short-circuit to ground.

KROHNE Group – Complete News

Got it, continue to print. F sensor electronic Error, operational fault in sensor electronics. Unscrew the cover of the electronic compartment and remove the two metal clips to the left and right of the display using a screwdriver or similar tool.

  R KELLY SOULACOASTER PDF

Improper installation and operation of the devices systems will cause the warranty to be void. Make sure not to damage the stranded drain wires 1, 20, Activation takes place only ofc setting mode is exited! It is permissible to lay the signal cable in water or in the ground.

All work on the electrical connections may only be carried out with the power disconnected. In case of DS cable, do not use connectors 20 and To Reset Counters C lrohne. Never mount the housing with the cable entries facing upwards. Signal converter monitors cable breaks and short circuits acc. C is set under “mode see above “, and this output is a “current output”.

The flowmeters are tested for a vibration level in accordance with IEC English, Danish, Polish China: This offers great time and cost benefits with regard to procurement, engineering, operation and servicing. Existing hardware cannot be identified. When it comes to process measurement, our level of expertise is unique, not just in standard applications but also for those challenges that demand customized solutions.

The following requirements regarding the electrical values of the signal cable must be observed: Process and ambient temperature too high. Be sure not to damage the stranded drain wire 1. Construction of signal cable B 1 Stranded drain wire for the inner shield 101. F hardware detection Existing hardware cannot be identified.

Krohme must be carried out in accordance with legislation applicable in your country.

Signaling of the active range possible via a status output adjustable. The field current cable is not part of the scope of delivery. A non-shielded three-wire copper cable is sufficient igc the field current cable.